Click to enlarge |
Rate:
- 84 bpm
- Regular
- Sinus rhythm - P waves best seen in leads V2-3
- Normal
- QRS - Normal
- Significant baseline artefact
- ST elevation leads V1-3
Interpretation:
- Baseline artefact
- Secondary to pre-existing Parkinson's disease
Troubleshooting artefact
There are multiple factors that can generate artifact during the ECG recording / printing process including:
- Patient factors
- Implanted devices
- PPM - Pacing Spikes
- Nerve stimulators
- Movement disorders e.g. Parkinson's, essential tremor
- Rigors
- Muscle tremor / activity can be related to pain, hypothermia etc.
- CPR
- Habitus - very thin or obese patients, paediatric patients
- Poor pad contact due to hair
- Poor pad contact due to moisture - sweating, immersion etc.
- Lead factors
- Poor pad contact
- Pad misplacement
- Damaged lead or lead connection
- ECG Machine Factors
- Inappropriate settings of gain or filtering modes
- Electrical artifact from power supply
- Printing factors
- Low ink
- Print head tracking
- Paper alignment
- External factors
- Interference from surrounding electronic devices. Relatively rare using newer ECG machines and environmental interference is usually compounded by incorrect machine filter settings.
An awareness of these factors and a systemic approach to addressing each point in the recording process, from patient to printer, is important in addressing recording problems such as artifact and lead mal-positioning.